/posts/til/til-researchhow-to-turn-room-temperature-into-a-real-test-condition
How to Turn ‘Room Temperature’ Into a Real Test Condition
Why This Matters When ‘room temperature’ lives only in your head, every experiment you run at it is slightly different and you won’t know by how much.
Why This Matters
When ‘room temperature’ lives only in your head, every experiment you run at it is slightly different and you won’t know by how much. I ran into this while trying to compare two sets of measurements months apart and realized my so-called 25°C tests had quietly ranged from 22°C to 28°C depending on the day and the chamber. Locking down what test temperature actually means turns a vague environmental note into a reproducible condition you can trust, compare, and defend in front of someone skeptical and detail-oriented.
How Test Temperature Actually Gets Set
In any real experiment, test temperature is not the number you type into the chamber; it’s the temperature the device under test (DUT) actually sees once everything has settled. There are at least three layers: the controller’s setpoint, the local environment around the DUT (air, oil bath, fixture), and the DUT’s own internal temperature, which may be higher or lower depending on self-heating and thermal resistance.
Chambers and baths are designed to control the environment, not the DUT directly. Their controllers usually regulate based on a single sensor in the air stream or fluid. If your DUT is bulky, powered, or shielded by fixtures, its temperature can lag or offset from that control sensor by several degrees. The trade-off is speed versus fidelity: aggressive control gives fast setpoint changes but can overshoot; gentler control gives smoother behavior but slower stabilization. Either way, the display can look perfect while the DUT is still drifting.
A robust temperature setting method therefore treats the DUT (or a surrogate mass right next to it) as the reference. You pick a nominal temperature (say, 25°C), then characterize what chamber setpoint and soak time are required for the DUT to sit within a tight band (for example, 24.5–25.5°C) and stay there. Once you know that mapping—setpoint, time, and load configuration—you can reproduce “25°C” as a physical condition, not just a number in a log.
The subtle but important step is to define what counts as ‘at temperature’ before you start collecting data: a specific sensor location, a tolerance band, and a minimum stable duration. That definition becomes part of your protocol. Without it, “room temperature” quietly expands to mean “whatever the lab felt like that day,” and every comparison you make across time or between setups is built on shifting sand.
💡 Did you know: Most cheap digital thermometers are factory-calibrated near room temperature, so they can be off by several degrees at oven or freezer temperatures unless you characterize them yourself.
Watching a DUT Reach ‘Test Temperature’
Setup
-----
• Chamber setpoint: 40.0 °C
• Ambient lab: 22.0 °C
• DUT: small PCB with sensor near hottest IC
• Sensors:
- T_chamber: controller sensor (air)
- T_dut: thermocouple taped to IC package
Timeline
--------
00:00 Power on chamber, set to 40 °C
00:10 T_chamber = 30 °C | T_dut = 24 °C (air heats quickly, DUT lags)
00:25 T_chamber = 39 °C | T_dut = 34 °C (controller nearly at setpoint)
00:35 T_chamber = 40 °C | T_dut = 37 °C (display looks 'ready')
00:50 T_chamber = 40 °C | T_dut = 39.5 °C (DUT still creeping up)
01:05 T_chamber = 40 °C | T_dut = 40.1 °C (both stable within ±0.2 °C)
Protocol decision
-----------------
Define 'at 40 °C test temperature' as:
• Condition: T_dut within 40.0 ± 0.5 °C
• Stability: drift < 0.2 °C over 10 minutes
• Start measurements no earlier than t = 60 minutes in this configuration.
Result
------
If you had started your test at 00:35 when the chamber first read 40 °C,
your DUT would actually have been at ~37 °C—off by 3 degrees and still rising.
The Insight
Test temperature is the steady-state temperature of the DUT, not the number you type into the chamber, so you have to define and verify it explicitly or your “25°C” today won’t match your “25°C” tomorrow.
🧠 Bonus: Many environmental test standards (like IEC 60068) allow ±2–3°C tolerance around the setpoint, which means two labs both claiming “25°C” can legally be several degrees apart unless you report the actual measured range.
Gotchas
- Dialing in temperature by the chamber’s setpoint instead of the DUT’s actual temperature → you think you’re testing at 25°C, but the part is still warming or cooling, so your data quietly mixes transient and steady-state behavior.
- Recording only a single nominal temperature (e.g., “tested at 40°C”) → later you can’t tell if a 2–3°C drift during the run explains a suspicious shift in results.
- Ignoring sensor placement and thermal mass → a big metal fixture can lag the air temperature by 10+ minutes, so starting measurements “when the display looks right” bakes in systematic bias.
- Mixing runs from different days that supposedly share a temperature setpoint → small seasonal or HVAC changes mean those 25°C runs might actually be 22°C vs 27°C, hiding real trends behind environmental noise.
Takeaways
- Define “at temperature” in advance using a specific sensor location, tolerance band, and stability criterion, instead of relying on whatever the chamber display says.
- Characterize the relationship between chamber setpoint and DUT temperature for your specific load and fixture, then reuse that mapping in future tests.
- Log both the nominal target temperature and the actual measured range during the run so you can later explain odd results or compare across experiments.
- Include soak time as a first-class parameter in your test plan; don’t start measurements until the DUT, not just the air, has demonstrably stabilized.
- When comparing results across days or labs, treat “25°C” as a hypothesis to verify with actual measurements, not a guarantee baked into the label.
🔥 One more thing: Thermal cameras often report apparent temperature assuming a default emissivity of 0.95; a shiny metal part can read 20–30°C off unless you correct emissivity or add a bit of matte tape as a reference patch.
References
- IEC 60068-3-5: Supporting documentation and guidance – Confirmation of the performance of temperature chambers (documentation)
- GUM: Guide to the Expression of Uncertainty in Measurement (book)
- Good Practice Guide No. 11: The Calibration of Temperature Block Calibrators (documentation)